A new intracavitary probe for detecting the site of origin of ectopic ventricular beats during one cardiac cycle.

B Taccardi, G Arisi, E Macchi, S Baruffi, S Spaggiari - Circulation, 1987 - Am Heart Assoc
B Taccardi, G Arisi, E Macchi, S Baruffi, S Spaggiari
Circulation, 1987Am Heart Assoc
An olive-shaped probe (25 X 12 mm) with 41 evenly distributed recording electrodes on its
surface was introduced into the left ventricles of seven open-chest dogs via the left atrium. In
two other dogs a cylindrical probe (40 X 3 mm) was used. Electrical stimuli were delivered at
66 endocardial, midwall, or epicardial sites in the left and right ventricular walls and the
septum. Mechanical stimuli were also applied at various epicardial sites. On-line mapping of
equipotential contour lines on the surface of the probe invariably revealed a clear-cut …
An olive-shaped probe (25 X 12 mm) with 41 evenly distributed recording electrodes on its surface was introduced into the left ventricles of seven open-chest dogs via the left atrium. In two other dogs a cylindrical probe (40 X 3 mm) was used. Electrical stimuli were delivered at 66 endocardial, midwall, or epicardial sites in the left and right ventricular walls and the septum. Mechanical stimuli were also applied at various epicardial sites. On-line mapping of equipotential contour lines on the surface of the probe invariably revealed a clear-cut potential minimum on the electrode that faced the pacing site. Time of appearance of potential minimum was 3 to 5 msec after endocardial stimuli, 10 to 25 msec for midwall and epicardial pacing, and 30 msec or more for right ventricular stimulation. Simultaneous stimulation at two sites 1.2 cm apart gave rise to two separate minima on the maps. "Pseudoisochrones" derived from electrograms recorded by the new probe were slightly less accurate in indicating the site of origin of extrasystoles. We conclude that equipotential and "isochrone" contour maps recorded from an array of semidirect electrodes, regularly distributed on the surface of an intraventricular probe, provide information on the site of origin (location and intramural depth) of ectopic paced beats in a normal dog heart.
Am Heart Assoc