Quality assurance for polychromatic flow cytometry using a suite of calibration beads

SP Perfetto, D Ambrozak, R Nguyen… - Nature protocols, 2012 - nature.com
SP Perfetto, D Ambrozak, R Nguyen, PK Chattopadhyay, M Roederer
Nature protocols, 2012nature.com
The quality assurance program presented here provides a means to maximize and maintain
the performance of individual flow cytometers in a facility. To optimize performance, we
recommend performing all three steps (optimization, calibration and standardization) in this
program when a new flow cytometer is installed or whenever the flow cytometer's optical
path is altered (eg, lasers, filters or detectors are replaced). The complete process requires 3–
4 h. On a more frequent basis, only a subset of these procedures need to be performed as a …
Abstract
The quality assurance program presented here provides a means to maximize and maintain the performance of individual flow cytometers in a facility. To optimize performance, we recommend performing all three steps (optimization, calibration and standardization) in this program when a new flow cytometer is installed or whenever the flow cytometer's optical path is altered (e.g., lasers, filters or detectors are replaced). The complete process requires 3–4 h. On a more frequent basis, only a subset of these procedures need to be performed as a part of daily maintenance routines. The data generated can be tracked to monitor the instrument and determine whether service is needed. In addition, the data can provide a metric for whether repairs and upgrades have improved or harmed performance, and for future instrument-to-instrument comparisons. In sum, the procedures presented here represent an updated framework for optimizing, calibrating and standardizing a flow cytometer for daily use.
nature.com